microAeth® / AL30 New

Tech specs

Measurement Method
Real-time Aethalometer® method by measuring the rate of change in absorption of transmitted light due to continuous collection of aerosol deposit on filter. Measurement at 880 nm interpreted as concentration of Black Carbon (‘BC’).
Measurement Wavelength
880 nm
Timebases (User setting)
1, 5, 10, 30, 60, or 300 seconds
Flow Rates (User setting)
Internal pump provides 25, 50, 75, 100, 125, 150, 170, 200, 225, or 250 ml/min
Measurement Range
0-1 mg BC/m3, filter life time dependent on concentration and flow rate setting:
  • avg. 5 μg BC/m3 for 24 hours @ 100 ml/min
  • avg. 100 μg BC/m3 for 3 hours @ 50 ml/min
  • avg. 1 mg BC/m3 for 15 minutes @ 50 ml/min
Measurement Resolution
0.001 μg BC/m3
Limit of Detection
0.010 µg BC/m3, 5 minute timebase, 250 ml/min flow rate <0.003 µg BC/m3, 1 hour average from 5 min timebase, 250 ml/min flow rate
Pump
Internal micro rotary vane pump
Flow Control
Internal mass flowmeter with closed-loop control
Filter Material / Capacity
Polytetrafluoroethylene (PTFE) filter strip material with 1 sampling location per filter
Sampling
3 mm spot created on filter strip containing insert of PTFE filter material. PM2.5 size selective inlet available.
Environmental Sensors
Temperature, Relative Humidity, Altimeter/Barometer, Accelerometer
Dimensions and Weight
Length: 98.5mm (3.878 inches)
Width: 61mm (2.401 inches)
Depth: 24.5mm (0.964 inch)
Weight: 162 grams (5.71 ounces)
Memory
8 MB internal flash memory, providing data storage up to:
  • 30 days operating on 5 minute timebase
  • 7 days operating on 1 minute timebase
Wireless Communications / Data Output
Data Stored in local memory.
Direct local data download over WiFi interface.
WiFi transmission to cloud data management system. Real-time and bulk data transfer when network is available 802.11 b/g/n Wi-Fi
Connections
USB-C (Power)
3.3V TTL Serial (Flow Calibration Communication)
Aerosol sample inlet and outlet ports
Total Run Time (Single battery charge)
Up to 24 hours @ 5 minute timebase, 50 ml/min flow rate. Run time may vary due to settings and PM concentrations.
Battery
Internal 3.7V 2200 mAh (8.14 Wh) 1 cell rechargeable lithium-ion battery.
Charging Time
Approximately 2.2 hours to full charge (using USB-C AC adapter, instrument turned off).
Power Supply Adapter
Input: 100~240 VAC 50/60 Hz 0.4 A
Output: 5VDC / 2A, USB-C
Operation Environment
5 ~ 40 °C operating, non-condensing.
Included
  • microAeth AL30
  • 1 Pack of 30 Filter Strips
  • 1 USB-C power cable
  • 1 territory-specific USB-C AC wall adapter
  • 1 one meter sampling hose with swivel tube connector
  • 1 Lapel clip for sampling hose
  • Manual available for download via AethLabs website
Updated August 2023, Revision 02. Specifications are subject to change without notice.