We've released a new version of the microAethCOM PC software (v2.1.0.1) and device firmware (PS706) which will provide several new options and improvements related to the operation microAeth AE51.
This software update to microAethCOM v2.1.0.1 is recommended for microAeth AE51 devices running the included firmware PS706 or the older firmware PS605. While we recommend updating the firmware on all AE51 devices to PS706, you can still maintain PS605 firmware on older devices, if you would like.
Improvements in microAethCOM v2.1.0.1
- Support for microAeth Model AE51 firmware PS706 (see improvements in PS706 below). AethLabs recommends updating your instrument firmware to PS706 for new features and improved performance.
- Data files can now be downloaded in either .DAT or .CSV filetypes to a disk location of the user's choosing.
- Instrument status codes are now displayed in bottom left hand corner of the main screen in human readable text if detected.
- Improved USB communications when used in conjunction with firmware PS706. Instrument startup and shutdown are detected in the microAethCOM console when used with firmware PS706.
- User interface has been simplified with instrument information shown on main page along with real-time graph.
- Data log is automatically deleted when an instrument is disconnected.
- Auto startup and shutdown options are no longer included in the microAethCOM application.
Improvements in firmware version PS706
- Simple shutdown mode. Hold the power button for three seconds to shut the instrument down. Secure shutdown mode is the timed sequence shutdown that uses multiple button presses to 'encode' the shutdown process. USB only mode can be used to lock the instrument from being shutdown.
- Improved USB communications in conjunction with PC software v2.1.0.1.
- Added 10s and 30s timebases.
- Improved measurement stability and sensitivity and flow control stability.
- Reintroduced the 200 ml/min flow rate setting
- Fixed flash full bug that causes instrument to not startup when memory is full.
- Improved flow calibration. Better measurement stability during the flow calibration process especially for the 25 ml/min datapoint. (Requires updated flow calibration software included with this install package.)
- Firmware PS706 can be installed on the AE51 once the microAethCOM v2.1 application is installed - just go to the 'Options' menu and select Upgrade Firmware. If you have an older microAeth AE51 that was manufactured by Magee Scientific with a serial number below 200, please contact AethLabs before attempting the upgrade process.
You must first either log in to or sign up for a free account at the AethLabs website to download the software. Once you are logged in, you can get more information and download the release here.
Post date: November 27, 2013 - 10:00pm