Measurement Method
Real-time Aethalometer® method, 5 wavelength absorption analysis by measuring the rate of change of transmitted light due to continuous particle deposition on filter. Measurement at 880 nm interpreted as concentration of Black Carbon (‘BC’). Measurement at 375 nm interpreted as Ultraviolet Particulate Matter (‘UVPM’) indicative of woodsmoke, tobacco, and biomass burning.
Measurement Wavelengths
880 nm, 625 nm, 528 nm, 470 nm, 375 nm
DualSpot® Loading Compensation
Real-time analysis by measuring the rate of change in absorption of transmitted light due to the continuous collection of aerosol on filter. Simultaneous collection on two spots in parallel at different flow rates.
Timebases
1, 5, 10, 30, 60, or 300 seconds
Flow Rates
Internal long life brushless diaphragm pump provides 100, 125, 150, 170, 200, 225, 250, 275, or 300 ml/min.
Measurement Range
Per sampling location, 0-1 mg BC/m3, filter sampling location lifetime dependent on concentration and flow rate setting, decreasing proportionally with lowest wavelength optical source enabled:
IR only mode, average 5 μg BC/m3 for 24 hours at 100 ml/min
IR only mode, average 100 μg BC/m3 for 3 hours at 50 ml/min
IR only mode, average 1 mg BC/m3 for 15 minutes at 50 ml/min
Measurement Resolution
0.001 μg BC/m3
Limit of Detection
30 ng BC/m3, 5 min timebase., 150 ml/min flow rate, SingleSpot
Pump Options
Internal long life brushless diaphragm pump
Flow Control
Internal mass flowmeters with closed-loop control
Filter Material / Capacity
MA600/MA350/MA300 Filter Tape Cartridge with Polytetrafluoroethylene (PTFE) material (85 sampling locations)
Sampling
3 mm diameter spot(s) created on filter tape. User selectable DualSpot ® or SingleSpot mode.
Aerosol Conditioning
Optional Internal Nafion Aerosol Dryer - uses internal vacuum pump in reflux configuration to dry inlet sample air stream
Environmental Sensors
Relative Humidity, Temperature, Altimeter/Barometer
Dimensions
3U 19 inch rack mount enclosure, W: 482.6 mm (19.0 in), H: 133.35 mm (5.25 in), D: 203.2 mm (8.0 in)
Weight
5 kilograms (11.02 pounds)
Memory
16 GB internal flash memory, providing storage for 31,250,000 data lines; 1 second timebase: 361 days of data.
On-board Interface
Low power screen, 3 buttons
Date/Time Format
ISO 8601 with manual computer synchronization
Connections
DB9 RS232 connector, 3.3V TTL Serial (Flow Calibration Communication), USB 2.0, M8 DC Power input connector, Quick Connect 6.35 mm (0.25 in) O.D. Tubing Aerosol sample inlet and outlet ports
USB Communications / Client Application
USB connectivity to cross-platform microAeth® Manager software available on macOS® and Windows®. microAeth Manager software is included and facilitates settings configuration and data download.
Serial Communication
DB9 connector with RS232 communication for streaming data and polling protocols to request data, modify settings and control. Command line interface (CLI) polling protocols: AethLabs protocol and Bayern-Hessen protocol., 3.3V TTL serial connectivity for uploading new instrument firmware, flow calibration.
Power
M8 12V Power In Power Supply Adapter: Input: 100~240 VAC 50/60Hz 0.4A, Output: 12VDC / 2A, with territory-specific Type A, C, G, or I plug
Operating Environment
5 ~ 40 °C operating, non-condensing.
Included
  • microAeth MA600
  • 1 MA600/MA350/MA300 Filter Tape Cartridge
  • M8 12V AC adapter with territory-specific plug
  • USB communication cable
  • Cross-platform microAeth® Manager software and manual available for download via AethLabs website
Accessories & Consumables
  • MA600/MA350/MA300 Filter Tape Cartridge
  • MA Series Flow Calibration Kit
  • microCyclone 170 PM2.5 Size-selective Inlet
  • Serial to USB converter cable
  • DB9 RS232 Serial to bare leads cable
  • Internal Aerosol Dryer
  • Inlet Protection Kit